• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG MININAV VALVE 10; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG MININAV VALVE 10; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FV660T
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 08/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing and quality control data the mininav® was manufactured by a qualified employee in november 2016.Deviations during assembly did not occur.The valve was sterilized by miethke and released for shipment after final inspection.The mininav® has a normal pressure range of 10 cmh2o.The valve was inspected as article fv660t.All parameters have been inspected and approved during the manufacturing process.All parameters have been assessed as meeting specifications.Visual inspection: no significant deformations or damage of the valve were detected during the visual inspection.Permeability test: a permeability test has shown that the valve is permeable.Computer controlled test: to investigate the claim of over/under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the valve is not operating within the acceptable tolerance in horizontal position.Internal inspection: after dismantling of the valve, some deposits were found in mininav® result: based on our investigation, we confirm that the valve shows a slower flow of liquid at the time of our investigation.This is likely due to the deposits observed inside the valves.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a mininav (part # fv660t) was implanted during a procedure performed on (b)(6) 2021.According to the complainant, the patient underwent a revision procedure on (b)(6) 2022.The reason the revision was performed was not provided.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 1 year, height: 75 centimeters (cm), weight: 10 kilograms (kg), gender: male.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MININAV VALVE 10
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15525462
Report Number3004721439-2022-00299
Device Sequence Number12276055
Product Code JXG
UDI-Device Identifier04041906133085
UDI-Public4041906133085
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K110206
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/07/2021
Device Model NumberFV660T
Device Catalogue NumberFV660T
Device Lot Number20032186
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2022
Initial Report FDA Received Date10/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/11/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age1 YR
Patient SexMale
Patient Weight10 KG
-
-