STRYKER TRAUMA KIEL LAG SCREWDRIVER GAMMA3® 380X110MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 1320-0200
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Insufficient Information (4580)
Date of Event 09/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that one of the teeth at the end of the lag screwdriver came off during the procedure.
 
Additional Manufacturer Narrative
Correction - d4 (lot#), h6 (clinical signs code, health impact code, component code).The reported event could be confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the received lag screwdriver shows traces of intense use identified by the appearance of scratches & dents on the surface.One of the four pegs was found to be broken and was not returned for evaluation.All the remaining pegs showed signs of deformation.Most likely the breakage occurred due to torsional overload.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the above investigation and available information, the root cause was attributed to user related issue.The breakage & deformation in pegs indicates towards torsional overloading.Moreover, the device was manufactured in 2011, so it can be said that it has performed its function in the previous surgeries as intended without any reported failure.Since it has been long in use, thus a normal weakened structural integrity due to repeated reprocessing cycles is considered possible.If any further information is provided, the complaint report will be updated.
 
Event or Problem Description
It was reported that one of the teeth at the end of the lag screwdriver came off during the procedure.The surgeon attempted to retrieve the fragment but it was unsuccessful.The procedure was completed successfully, but a fragment from the ¿screwdriver¿ device had to be left in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG SCREWDRIVER GAMMA3® 380X110MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15526176
Report Number0009610622-2022-00449
Device Sequence Number2674692
Product Code LXH
UDI-Device Identifier04546540377807
UDI-Public04546540377807
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1320-0200
Device Catalogue Number13200200
Device Lot NumberK949634
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/07/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date10/02/2022
Supplement Report FDA Received Date12/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/04/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-