• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSTEOMED, LLC INSTAFIX IMPLANT KIT, 10X10MM; STAPLE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSTEOMED, LLC INSTAFIX IMPLANT KIT, 10X10MM; STAPLE, FIXATION, BONE Back to Search Results
Model Number 400-3703-SP
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is currently pending, and a follow up report will be submitted upon completion of the investigation.
 
Event or Problem Description
It was reported the patient underwent the initial implant procedure for hallux interphalangeal joint fusion on (b)(6) 2022.The surgeon reported the patient had a non-union and one of the two 10 x 10 instafix staples had broken at the corner (event date unknown).On (b)(6) 2022, the surgeon explanted both instafix staples, debrided, fenestrated, added dbm putty, and placed two 15 x 10mm instafix staples.It was indicated the explanted instafix staples were not available to be returned for evaluation.This report is related to report number 2027754-2022-00050 for the other explanted instafix staple.
 
Additional Manufacturer Narrative
Sections d4.Udi and h4.Device manufacture date completed as review of manufacturing and inspection records has been completed.The device was not returned for evaluation.The device history record (dhr) was reviewed for lot 20622992102, and no issues were identified in the dhr review.The instafix kits from lot 20622992102 were issued staples from bulk lot 1161093.All inspections passed and no issues were identified with this bulk lot.A two year review (november 2020 - october 2022) of the complaint database found eight (8) complaints related to the instafix kits.This is the only reported complaint for a broken staple/non-union of bone.Review of the instructions for use (ifu) revealed the warning section states: "the devices can break when subjected to increased loading associated with nonunion or delayed union.Internal fixation devices are load sharing devices that hold a fracture in alignment until healing occurs.If healing is delayed, or does not occur, the implant can be expected to break, deform or fail.Loads produced by weight bearing, and activity levels may dictate the longevity of the implant." follow up with the field revealed "the surgeon did not indicate that he had any theories on what caused the non-union".Based on the information received and the investigation performed, the root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSTAFIX IMPLANT KIT, 10X10MM
Common Device Name
STAPLE, FIXATION, BONE
Manufacturer (Section D)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer (Section G)
OSTEOMED, LLC
3885 arapaho rd
addison TX 75001
Manufacturer Contact
ellie wood
3885 arapaho rd
addison, TX 75001
9726774600
MDR Report Key15526250
Report Number2027754-2022-00049
Device Sequence Number2674693
Product Code JDR
UDI-Device Identifier00842528104170
UDI-Public(01)00842528104170(17)000101(10)20622992102
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K141550
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number400-3703-SP
Device Catalogue Number400-3703-SP
Device Lot Number20622992102
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2022
Supplement Date Received by Manufacturer10/24/2022
Initial Report FDA Received Date10/02/2022
Supplement Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-