| Model Number |
B3400060 |
| Medical Device Problem Codes |
Human-Device Interface Problem (2949); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during a primary implant today, the setscrew from two sensight extensions dislodged during tunneling and remained in the extension track (approximately lower neck, subcutaneous).Spare setscrews were utilized to lock the lead into the extension.The surgeon opted to leave the dislodged setscrews in the body.If the patient cannot have an mri, it would require a further operation to attempt to explant the setscrews.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id: b3400060m, serial#: (b)(6), product type: extension.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received: it was reported that the scrub nurse was used to using boston scientific extensions where they back the screw our before insertion of the dbs lead.Unfortunately the nurse also backed the screws out for our extensions, enough so they just stayed in place until they were pulled through the neck where they became dislodged.
|
| |
|
Search Alerts/Recalls
|