• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. OXFORD TRIAL BRNG LRG SZ 4/5; KNEE INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXFORD TRIAL BRNG LRG SZ 4/5; KNEE INSTRUMENT Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported, that the trial bearing cracked where it screws into the handle.No consequences or impact to the patient.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual inspection of the returned instrument shows that the 4mm trial bearing has fractured at the point where the handle fits the trial bearing.General wear & tear is evident and is expected due to repeated use during its time in the field (approximately 6 years and 10 months).Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.Medical records were not provided.The root cause of the reported issue is attributed to expected wear and tear from repeated use and reprocessing during its time in the field (6 years and 10 months).If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXFORD TRIAL BRNG LRG SZ 4/5
Common Device Name
KNEE INSTRUMENT
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key15526863
Report Number3002806535-2022-00419
Device Sequence Number6056375
Product Code HRY
UDI-Device Identifier05019279483891
UDI-Public(01)05019279483891(11)160103(10)ZB111101
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number32-420382
Device Lot NumberZB111101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date10/03/2022
Supplement Report FDA Received Date10/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/03/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-