• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VICRYL SUTURE 27" (70CM) 0 VIO; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VICRYL SUTURE 27" (70CM) 0 VIO; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number J318H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #pc-(b)(4).Component code: (b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Could you please clarify if just one device present the two issues (breakage suture & breakage needle) or one device present breakage suture and other present breakage needle? please confirm.Could you please clarify when did the issues occurred, (removal from package /during passage through tissue/ during tying / post-op)? please confirm.Event related to mw # 2210968-2022-08131.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6)2022 and suture was used.The sutures broke during use and the needle also broke.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 11/29/2022.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection evaluation was conducted on the returned device.Visual analysis of the returned sample determined that one empty folder and one needle-suture piece that pertains to product code j318h were returned to ethicon for analysis.In order to evaluate the conditions of the returned sample, the suture piece was observed with body fluids and the end was broken possibly caused by a surgical instrument.The product code j318h contains absorbable suture and as the sample was received open, the time of exposure of suture to the environment could not be determined and no functional test can be performed due to sample condition.The condition of the sample received indicates improper handling of the device.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2022-08130 & 2210968-2022-08131.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VICRYL SUTURE 27" (70CM) 0 VIO
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-BRAZIL
rodovia presidente
dutra, km 154
sao paolo 12240 -908
BR   12240-908
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15527208
Report Number2210968-2022-08130
Device Sequence Number14099261
Product Code GAM
UDI-Device Identifier17891338030612
UDI-Public17891338030612
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2024
Device Model NumberJ318H
Device Catalogue NumberJ318H
Device Lot NumberAM3863
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2022
Supplement Date Received by Manufacturer11/07/2022
Initial Report FDA Received Date10/03/2022
Supplement Report FDA Received Date11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
-
-