TELEFLEX MEDICAL EZ-BLOCKER KIT; EZ BLOCKER Back to Search Results
Model Number IPN913391
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The complaint reported that: "we inflated the green balloon to isolate the right mainstem.During the case, right upper lobe begin to inflate.".
 
Event or Problem Description
The complaint was reported as: "we inflated the green balloon to isolate the right mainstem.During the case, right upper lobe begin to inflate.Additional air was injected.Then same issue occurred.Then after a while, total tidal volume on the vent began to decrease then pt's saturation began to decrease.The right (green) balloon was deflated and the position of the balloon was examined bronchoscopically, which was fine.This occurred at at least a couple of times.The anesthesiologist noticed that the left (blue) balloon kept inflating.So, what she did was to deflate the left balloon (blue) and inject 1 ml of air into the right (green).This solved the issue.On examination of the device after the case, i was able to inflate the blue balloon by compressing the green balloon.So my conclusion is that there was a tiny communication between the two channels." no patient injury or consequence reported.The patient condition was reported as "fine".
 
Additional Manufacturer Narrative
Qn#(b)(4).Additional information was received regarding the patient desaturation and the sales rep has reported "the lowest was 94%, but she was in the high 90s most of the case." the actual sample was returned and sent to the manufacturing site for evaluation.The manufacturing site reports that a visual exam was performed and no defects were observed.It was also reported that functional testing was performed and "15ml of the air was let in the balloon on yellow extrusion through yellow pilot valve and after few minutes the blue pilot valve and the balloon on blue extrusion and started slowly inflating." a device history record review was performed and no relevant findings were identified.The manufacturing site reports that the complaint was confirmed based on the functional inspection.The root cause is listed as supplier related.A capa was opened to address this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ-BLOCKER KIT
Common Device Name
EZ BLOCKER
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15527273
Report Number3006425876-2022-00844
Device Sequence Number11431550
Product Code CBI
UDI-Device Identifier10801902142372
UDI-Public10801902142372
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K121462
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/26/2023
Device Model NumberIPN913391
Device Catalogue NumberMG-02770-002
Device Lot Number71F21J2659
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2022
Supplement Date Received by Manufacturer10/24/2022
Initial Report FDA Received Date10/03/2022
Supplement Report FDA Received Date10/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale
Patient Weight104 KG
-
-