C.R. BARD, INC. (COVINGTON) -1018233 SILICONE FLAT DRAIN; HUBLESS SILICONE FLAT DRAIN Back to Search Results
Model Number 0070370
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the drain bag had fractured.It was also noted that the patient had underwent a kidney transplant during which a 10fr flat drain bag was placed and when the physician attempted to remove the drain it was noticed from the appearance that drain had fractures and the pieces were also retained.Later patient was taken to the operating room for the removal of the retained drain fragments.
 
Event or Problem Description
It was reported that the drain bag had fractured.It was also noted that the patient had underwent a kidney transplant during which a 10fr flat drain bag was placed and when the physician attempted to remove the drain it was noticed from the appearance that drain had fractures and the pieces were also retained.Later patient was taken to the operating room for the removal of the retained drain fragments.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.One sample were confirmed to exhibit the reported failure.The device had not met specifications.The product was used for patient treatment.Visual evaluation of the returned photo sample noted one opened (without original packaging), used hubless flat drain.Visual inspection of the photo sample noted a piece broken off from the hubless flat drain.This does not meet specification, which states "no damaged or missing components".A potential root cause for this event could be ¿drain was manufactured incorrectly (dimensions or material out spec, curing time not appropriate, etc.)".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "reliavac® closed wound suction system with hubless silicone drains instructions for use indications for use: wound drains are used to remove exudates from wound sites.Contraindications: do not use for chest drainage.Warning: do not bypass/inactivate anti-reflux valve.Sterilized by ethylene oxide caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Important: a.Check for fluid entering reliavac® evacuator.Lack of flow may indicate: all exudate has been removed.Wound drain is clogged and may require irrigation and aspiration (consult physician).Auxiliary wall suction pressure is above 210mm hg.Deflated balloon: check all connections for air leak and wound tube perforations for exposure above the skin.If still deflated, replace evacuator.B.When not using auxiliary suction during surgical wound closure, several activations of the reliavac® evacuator may be required to establish suction because of: air entering partially closed wound.An operative air pocket.C.Insert safety pin into hole in collar to attach evacuator to patient's clothing or bed linen.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.To avoid the possibility of a hematoma due to wound evacuation, the instructions for use should be carefully followed.To avoid the possibility of drain damage or breakage: additional perforations should not be made in the drains.Avoid suturing through drains.Drains should lie flat and in line with the skin exit areas.Particular care should be taken to avoid any obstacles to the drain exit path.Drains should be checked during closure for free motion to avoid possibility of breakage.Drain removal should be done gently by hand.They should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.Warning: surgical removal may be necessary if drain is difficult to remove or breaks.A.Use with single flat drain - 1.Place perforated wound drain within critical fluid collection area of wound.2.Draw drain tube through skin/stab wound incision until flat portion of drain is seated appropriately.3.Trim drain tube to desired length and attach to blue adapter.Connect other end of blue adapter to y-connector.4.Insert connecting tube in reliavac® port a, up to indicator ring.B.Use with two flat drains - 1.Place perforated portion of wound drains within critical fluid collection areas of wound.2.Draw drain tubes through skin/stab wound incision until flat portion of the drains are seated appropriately.3.Trim drain tubes to desired length.4.Cut off plug from closed arm of y-connector and attach blue adapters.5.Attach drains to blue adapters.6.Insert connecting tube in reliavac® port a, up to indicator ring.Caution: punctures or additional perforations should not be made in the silicone wound drain.C.Attaching to auxiliary suction - 1.Insert suction adapter into port b.2.During auxiliary suction, balloon will inflate and exudate will flow over balloon surface from port a to port b.3.To discontinue auxiliary suction, remove suction adapter and close port b.Caution: do not use with wall suction in excess of 210mm hg.D.To establish suction - 1.Open port b.2.Pump bulb until balloon fills container.3.Close port b.Note: hissing sound is normal and stops when maximum suction pressure is reached.Possible reflux of fluid to the patient is reduced during reliavac® evacuator reactivation by a built-in anti-reflux valve in port a.E.To empty container - 1.Open port b.2.Invert unit.3.Pump bulb to empty quickly.F.To re-establish suction - 1.Repeat step "d" above.G.To read fluid volume - 1.Open port b.2.Allow balloon to deflate.3.Read and record volume.4.To reactivate, repeat step "d" above.Manufacturer: c.R.Bard, inc.Covington, ga 30014 usa 1-800-526-4455 www.Bardmedical.Com bard and reliavac are trademarks and/or registered trademarks of c.R.Bard, inc.© 2017 c.R.Bard, inc.All rights reserved." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILICONE FLAT DRAIN
Common Device Name
HUBLESS SILICONE FLAT DRAIN
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15527764
Report Number1018233-2022-07609
Device Sequence Number11409044
Product Code GBX
UDI-Device Identifier00801741090745
UDI-Public(01)00801741090745
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/19/2025
Device Model Number0070370
Device Catalogue Number0070370
Device Lot NumberNGFZ4211
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/23/2022
Supplement Date Received by Manufacturer02/24/2023
Initial Report FDA Received Date10/03/2022
Supplement Report FDA Received Date02/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-