DORNOCH DUO FLUID CART - REFURBISHED; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Codes |
Melted (1385); Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the power inlet module and power cord was melted, the fuses were shorted out, the fuse ports were blistered, and the cart would not power on.The power inlet module and power cord were replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the power cord melted.The event timing was outside of surgery.There was no harm or delay reported.Upon investigation, it was found that the power inlet module and power cord were burnt and melted, the fuses were shorted out, the fuse ports were blistered.No adverse events were reported as a result of this malfunction.
|
| |
|
Search Alerts/Recalls
|
|
|