| Model Number |
401744 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was selected for treatment.The orsiro was inserted into the patients body, not knowing that the stent was not on the balloon, but had fallen off on the back table before insertion.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Neither the complaint instrument nor the angiographic material was returned for analysis.Therefore, no technical investigation on the subject could be performed.The product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations, no manufacturing or material related root cause could be determined.It should be noted that the ifu advises the user to visually inspect the stent system prior to the procedure to ensure that the stent is centered between the two radiopaque markers on the balloon.
|
| |
|
Search Alerts/Recalls
|