| Model Number |
0684-00-0478-01 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter full name: (b)(6).Event site postal code: (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).Device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during intra-aortic balloon (iab) therapy, the patient was getting ready for an x-ray at the bed side.All of a sudden, the gas extension of the iab became filled with blood.The cs300 intra-aortic balloon pump (iabp) generated a blood detected alarm and the iab was removed.The patient was waiting for surgery the next day, but was in stable condition.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint #(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned with the membrane completely unfolded and blood on the interior and exterior of the catheter and between the catheter and the sheath.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and one leak was detected on the membrane approximately 1.5cm from the rear seal measuring 0.025cm in length.The reported problem was most likely triggered by a leak which was found on the membrane.Under magnification, a whitish patch was observed around the leak.This whitish patch is the typical appearance of an abrasion mark which is caused by calcified plaque in the aorta.A lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint #(b)(4).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|