APPLIED MEDICAL RESOURCES EB216, VOYANT MARYLAND FUSION 44CM, 6/BX; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number EB216
Medical Device Problem Code Self-Activation or Keying (1557)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Testing was performed on the returned unit, where it was observed that the device would unintentionally deliver rf energy, confirming the complainant¿s experience.Upon closer inspection, an internal fuse switch, was observed to be misaligned.Based on the condition of the returned unit and the description of the event, it is likely that the reported event was caused by the misaligned fuse switch.Based upon the initial description of the event, the event was not reportable as it is unlikely to cause or contribute to death or serious injury.However, based on the evaluation of the returned unit, applied medical determined that this event is reportable as unintended rf output could cause or contribute to death or serious injury.This report is a combined initial and follow-up report.
 
Event or Problem Description
Procedure performed: laparoscopic gastric sleeve.Event description: surgeon reports that fuse activation button would cause delayed and prolonged response during use.Energy delivery occurred a few seconds after pressing the fuse activation button in procedure.Upon releasing fuse activation button and removing device from tissue, energy was still being delivered to the device until an error code interrupted the seal cycle.A new device was opened to complete the case.No patient injury occurred.Product is available for return.Intervention: a new device was opened to complete the case.Patient status: no patient injury occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EB216, VOYANT MARYLAND FUSION 44CM, 6/BX
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key15530562
Report Number2027111-2022-00755
Device Sequence Number19459422
Product Code GEI
UDI-Device Identifier00607915139360
UDI-Public(01)00607915139360(17)250213(30)01(10)1444896
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K200598
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/13/2025
Device Model NumberEB216
Device Catalogue Number101475464
Device Lot Number1444896
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2022
Initial Report FDA Received Date10/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-