Catalog Number 07.01702.007
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Fall (1848)
Date of Event 08/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2022-00217 through 3012447612-2022-00219.
 
Event or Problem Description
It was reported that two virage screws loosened and one disassembled post-operatively after the patient may have fallen.A revision surgery was performed where the three screws were removed and replaced, and the construct was extended from levels c4 - t2 to c3 - t2.This is report three of three for this event.
 
Event or Problem Description
It was reported that two virage screws loosened and one disassembled post-operatively after the patient may have fallen.A revision surgery was performed where the three screws were removed and replaced, and the construct was extended from levels c4 - t2 to c3 - t2.This is report three of three for this event.
 
Additional Manufacturer Narrative
Corrections in d9 and h3.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation device inspection revealed that two pieces from lot abq had cracks in the weld on one side.The third piece from lot abq had both welds broken, causing the lower housing to break into its two pieces and allowing the screw components to disassemble.The weld penetration depth was confirmed to be greater than minimum specification.Potential cause the root cause was unable to be determined.This event could possibly be attributed to the patient's fall or other unknown patient and operational factors.Dhr review per dhr review, the parts were likely conforming when they left zimvie control.Device use and compatibility this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYAXIAL SCREW (DIA. 3.5 MM X 14 MM L)
Common Device Name
VIRAGE OCT SPINAL FIXATION SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15530758
Report Number3012447612-2022-00219
Device Sequence Number6055475
Product Code KWP
UDI-Device Identifier00889024320925
UDI-Public(01)00889024320925(10)ABQ
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133556
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07.01702.007
Device Lot NumberABQ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2022
Supplement Date Received by Manufacturer03/28/2023
Initial Report FDA Received Date10/03/2022
Supplement Report FDA Received Date03/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-