SYNTHES GMBH TFNA HELICAL BLADE PERF L90 TAN; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
|
Back to Search Results |
|
| Model Number |
04.038.390S |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
09/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional procode: hsb.Device available for evaluation: complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from japan reports an event as follows: it was reported that on (b)(6) 2021, the patient underwent orif for left femoral trochanteric fracture.After surgery, the patient was rehabilitated in another hospital and the bone union was confirmed in (b)(6) 2022.At the end of (b)(6) 2022, the patient fell and had pain.On (b)(6) 2022, the patient visited the hospital and it was confirmed that the blade had perforated the bone head.It was planned to remove the tfna in question and perform tha surgery on (b)(6) 2022.The surgeon commented that the blade perforated the bone head due to the patient fall, because the bone union was confirmed in april.It was possible that lsc (late segmental collapse) occurred.No further information is available.This report is for a tfna fenestrated helical blade 90mm - sterile.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Manufacturing location: elmira / packaged, sterilized and released by: monument manufacturing date: 15-mar-2021 expiration date: 01-mar-2031 part number: 04.038.390s, tfna fenestrated helical blade 90mm -sterile lot number: 94p6678 (sterile).Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|