| Model Number |
1012447-12 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported balloon rupture appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a heavily calcified, moderately tortuous lesion in the right coronary artery (rca).Rotablation was performed.The 2.5x12 mm nc trek balloon dilatation catheter (bdc) ruptured at 12 atmospheres on the first inflation due to the heavy calcification.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|