SYNTHES GMBH UNK - SCREWS: TRAUMA; SCREW, FIXATION, BONE Back to Search Results
Model Number 04.211.022
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Date of event: unknown date of breakage.Device available for evaluation: complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter name: initial reporter is j&j company representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from japan reports an event as follows: it was reported that on (b)(6) 2022, the patient underwent the open reduction internal fixation surgery for fracture of distal clavicle with 5 screws.The surgery was completed successfully.On (b)(6), it was confirmed that 5 distal screws were broken.It is unknown when the screws broke.The surgeon commented that the cause was unknown.The patient had dementia.Concomitant device reported: 2.7 ti va lcp clav pl lat/ cs1/ lft/ -s (04.112.610s, 640p853).This is report 6 of 9 for (b)(4).This report is 2.7mm ti cortex screw slf-tpng with t8 stardrive recess 20mm.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Part # 04.211.022ts, lot # 8l96329, manufacturing site: werk selzach logistik, release to warehouse date: 10 jan 2022, expiration date: 01 jan 2027, supplier: (b)(4).Non-sterile part # 04.211.022ts, non-sterile lot #546p509, manufacturing site: werk selzach logistik, release to warehouse date: 07 dec 2021, supplier: (b)(4).The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from the attachment.Visual analysis of the photo revealed that there was no damage or defects with the unk - screws: trauma.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the unk - screws: trauma was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
This report is for one (1) unknown screw this is report 2 of 5 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK - SCREWS: TRAUMA
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15532606
Report Number8030965-2022-07732
Device Sequence Number14766428
Product Code HWC
UDI-Device Identifier10886982119032
UDI-Public(01)10886982119032
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K100776
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.211.022
Device Catalogue Number04.211.022TS
Device Lot Number8L96329
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/08/2022
Supplement Date Received by Manufacturer10/06/2022
12/07/2022
Initial Report FDA Received Date10/03/2022
Supplement Report FDA Received Date10/28/2022
12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2.7 TI VA LCP CLAV PL LAT/ CS1/ LFT/ -S; CORTSCR ø2.7 SELF-TAP L20 TAN; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L20 TAN; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L20 TAN; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L20 TAN; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L22 TAN; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L22 TAN; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L22 TAN; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L24 TAN; VA LOCKSCR ø2.7 HE 2.4 SELF-TAP L24 TAN
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-