Model Number MCENDO550
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported to terumo cardiovascular that there was a scope issue.It is unknown when this event occurred, whether there was a delay in the procedure, whether the product was changed out, or if there was any effect on the patient or results of the surgery.Due to the unknown information for this event, it is being reported.Terumo continues to attempt to gain more information regarding this event from the user facility.
 
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.(b)(4).
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on october 3, 2022.Upon further investigation of the reported event, the following information is new and/or changed: g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information) h6 (identification of evaluation codes 4114, 3224, 3259, 19) type of investigation: 4114 - device not returned.Investigation findings #1: 3224 - optical problem identified.Investigation findings #2: 3259 - improper physical structure.Investigation conclusions: 19 - cause traced to use the affected sample was not returned; therefore, a direct investigation could not performed.Past product complaints were reviewed for similar issues and possible causes are most likely due to, the product was either dropped or inadvertently struck by an object, and excessive force was exerted on the endoscope shaft, that could deflect the shaft and lead to the alteration of the internal lens system.The most likely root cause is improper handling caused the internal lens to break.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE 5.5MM
Common Device Name
LAPAROSCOPE, GENERAL
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key15533017
Report Number1124841-2022-00156
Device Sequence Number5212396
Product Code GCJ
UDI-Device Identifier00699753018952
UDI-Public(01)00699753018952
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K092789
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberMCENDO550
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date10/03/2022
Supplement Report FDA Received Date12/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-