Model Number 1012447-20
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.The additional devices referenced in b5 are being filed under separate medwatch report numbers.
 
Event or Problem Description
It was reported that the procedure was to treat a bifurcation in the first diagonal artery.The 2.5x20mm nc trek rx balloon dilatation catheter (bdc) was prepared and flushed prior to use but was difficult to advance trapping the distal part of the hi-torque (ht) balance middleweight (bmw) universal ii guide wire.The guide wire had resistance upon removal with the balloon catheter and some force was applied breaking the proximal core of the guide wire but remaining intact.The guide wire and the balloon were simply removed together.Another guide wire and balloon were used to complete the procedure.Another balloon was difficult to advance over another ht bmw universal ii guide wire in the anterior descending artery to perform the final kissing balloon technique (kbt) requiring the guide wire to be changed.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Visual, dimensional and functional inspections were performed on the returned device.The reported difficulty advancing and removing the device over the guide wire was not confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported difficulty advancing and removing the device from the guide wire.Factors that may contribute to the difficult advancing and removing the bdc from the guide wire causing resistance between the devices may include, but not limited to, manufacturing damage, device placement technique, buildup of procedural contaminants in the guide wire lumen, guiding catheter support, inner diameter of guide wire lumen, outer diameter of the guide wire, condition of the guide wire, or damage to the distal shaft of the catheter.In this case, a conclusive cause for the reported difficulty could not be determined since the complaints could not be confirmed during return analysis.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a bifurcation in the first diagonal artery.The 2.5x20mm nc trek rx balloon dilatation catheter (bdc) was prepared and flushed prior to use but was difficult to advance trapping the distal part of the hi-torque (ht) balance middleweight (bmw) universal ii guide wire.The guide wire had resistance upon removal with the balloon catheter and some force was applied breaking the proximal core of the guide wire but remaining intact.The guide wire and the balloon were simply removed together.Another guide wire and balloon were used to complete the procedure.Another balloon was difficult to advance over another ht bmw universal ii guide wire in the anterior descending artery to perform the final kissing balloon technique (kbt) requiring the guide wire to be changed.There was no adverse patient effect and there was no clinically significant delay in the procedure.Subsequent to the initially filed report, the following information was provided: the ht bmw universal ii guide wire did not separate in the patient's anatomy and it was confirmed there was nothing left in the patient.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15533454
Report Number2024168-2022-10251
Device Sequence Number12720614
Product Code LOX
UDI-Device Identifier08717648151774
UDI-Public08717648151774
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model Number1012447-20
Device Catalogue Number1012447-20
Device Lot Number20223G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer11/28/2022
Initial Report FDA Received Date10/03/2022
Supplement Report FDA Received Date12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HT BMW UNIVERSAL GUIDE WIRE
Patient SexFemale
-
-