MAUDE Adverse Event Report: MENTOR TEXAS MENTOR CPX4 PLUS, SMOOTH, MEDIUM HEIGHT; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR CPX4 PLUS, SMOOTH, MEDIUM HEIGHT; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number SCPX117MH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: n/a.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a mentor cpx4 plus, smooth, medium height tissue expander being used in an unspecified breast implantation procedure was found to be deflated during the operation.No adverse event was experienced by the patient.The surgeon used backup devices, and there were no reported patient consequences or procedural delays.
 
Additional Manufacturer Narrative
Device evaluation summary: the product was returned to mentor for evaluation.Mentor then conducted a visual inspection of the returned device.During the visual analysis of the returned sample cpx4 plus sm te mh 350cc, a leakage was observed, caused by delamination dome.Based on the information currently available, a product issue was identified during the investigation of the product received.This product issue will be addressed through the mentor¿s quality system.In addition, a non conformance (nc) was opened to address this product issue.According to the manufacturing investigation, an additional analysis was performed on the device and injection site.The analysis confirmed the presence of the polyethylene that is removed from the surface of the washer before being used to assemble the injection site to the shell.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR CPX4 PLUS, SMOOTH, MEDIUM HEIGHT
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
michelleann garcia
3041 skyway circle north
irving, TX 75038
646591-7981
MDR Report Key15533682
Report Number1645337-2022-11634
Device Sequence Number8133181
Product Code LCJ
UDI-Device Identifier10081317028366
UDI-Public10081317028366
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K182335
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/20/2026
Device Model NumberSCPX117MH
Device Catalogue NumberSCPX117MH
Device Lot Number9708237
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/05/2022
Supplement Date Received by Manufacturer12/07/2022
Initial Report FDA Received Date10/03/2022
Supplement Report FDA Received Date12/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-