ABBOTT MEDICAL SJM RIGID SADDLE RING; RING, ANNULOPLASTY Back to Search Results
Model Number RIGID SADDLE RING
Medical Device Problem Code Lack of Effect (4065)
Health Effect - Clinical Code Mitral Valve Insufficiency/ Regurgitation (4451)
Date of Event 09/10/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2022, a 26mm rigid saddle ring was successfully implanted in a patient.The ring had been sized using a ring sizer set.However, post-implant, there was severe mitral regurgitation (mr) confirmed via transesophageal echocardiogram examination.The patient had been taken off bypass to check effectiveness of the ring, but was placed back on bypass after confirming severe mr.The ring was explanted and replaced intraoperatively with a 25mm sjm masters series valve with expanded cuff to complete the procedure.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects reported.The patient was stable and recovering at the time of report.The cause for the severe mitral regurgitation is unknown.
 
Additional Manufacturer Narrative
An event of explant due to severe mitral regurgitation was reported.The device met visual specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the reported event could not be conclusively determined.Images from field appeared to show a ring when attached and detached from holder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM RIGID SADDLE RING
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15533879
Report Number2135147-2022-01455
Device Sequence Number12076267
Product Code KRH
UDI-Device Identifier05414734009423
UDI-Public05414734009423
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K042734
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/27/2026
Device Model NumberRIGID SADDLE RING
Device Catalogue NumberRSAR-26
Device Lot Number8034259
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/12/2022
Supplement Date Received by Manufacturer01/03/2023
Initial Report FDA Received Date10/03/2022
Supplement Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-