• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VERITY ADX XL VDR; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VERITY ADX XL VDR; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5456
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
During the implant procedure, the set screw was unable to be tightened.The event was resolved by replacing the device.The patient was in stable condition.
 
Additional Manufacturer Narrative
The reported event of could not tighten the atrial lead in the pacemaker was confirmed.Analysis revealed that many incorrect torque wrench insertions were made by the user which damaged the septum punching out excessive pieces of the septum.These septum punch-outs landed in the setscrew inset filling it to the top.The user was incorrectly inserting the wrench and now if any correct wrench insertions were made, they were blocked by the punch-outs in the inset.In the user's attempts to engage the setscrew the user screwed the setscrew out of the connector block socket.The setscrew then fell loose and sideways across the block.When this happens, the wrench will only hit the sides of the setscrew and it is now impossible to engage and tighten.The problem was caused by the user¿s incorrect use of the torque wrench.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERITY ADX XL VDR
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15534547
Report Number2017865-2022-39671
Device Sequence Number12065224
Product Code DXY
UDI-Device Identifier05414734006910
UDI-Public05414734006910
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model Number5456
Device Catalogue Number5456
Device Lot NumberP000153373
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer11/22/2022
Initial Report FDA Received Date10/03/2022
Supplement Report FDA Received Date11/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RIGHT ATRIAL LEAD
Patient SexUnknown
-
-