• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 40MM/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 40MM/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 6260-9-240
Medical Device Problem Codes Corroded (1131); Material Erosion (1214)
Health Effect - Clinical Codes Fall (1848); Metal Related Pathology (4530); Swelling/ Edema (4577)
Date of Event 07/26/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his left hip on or about (b)(6), 2015 and was revised on (b)(6), 2022.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his left hip on or about (b)(6) 2015 and was revised on (b)(6) 2022.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.
 
Additional Manufacturer Narrative
Reported event: an event regarding corrosion, abnormal ion level involving a metal head was reported.The event of corrosion was confirmed.Abnormal ion level was not confirmed.Method & results: product evaluation and results: material, dimensional, functional and material analysis could not be performed as the device was not returned.Clinician review: a review of the provided medical information by a clinical consultant indicated: conclusion/assessment: a patient underwent a tha with a v40 cocr head 40mm and an accolade ii stem.Approximately 71/2 years later the patient underwent a revision for pain after having a fall and mri showing tearing of the abductors, a fluid collection, osteolysis.The preop laboratory studies did not show infection but had reported increase in co level (normal levels not provided).At the time of the revision there were signs of trunnionosis and metallosis.The pre and postoperative medical records, radiographs, and laboratory reports were not provided for review.Event confirmation: a revision surgery can be confirmed.Metalloisis and presumed trunnionosis can be confirmed.Increased metal levels were reported in the operative note, but the laboratory reports not provided.Device recall and injury cannot be confirmed.Root cause: the root cause of the revision surgery was pain after a fall, fracture of the trochanter, and what appeared to be trunnionosis with soft tissue metal response.The root cause of the trunnionois cannot be determined.The root cause of the other allegations cannot be determined as they cannot be confirmed.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of the provided medical information by a clinical consultant indicated: conclusion/assessment: a patient underwent a tha with a v40 cocr head 40mm and an accolade ii stem.Approximately 71/2 years later the patient underwent a revision for pain after having a fall and mri showing tearing of the abductors, a fluid collection, osteolysis.The preop laboratory studies did not show infection but had reported increase in co level (normal levels not provided).At the time of the revision there were signs of trunnionosis and metallosis.The pre and postoperative medical records, radiographs, and laboratory reports were not provided for review.Event confirmation: a revision surgery can be confirmed.Metalloisis and presumed trunnionosis can be confirmed.Increased metal levels were reported in the operative note, but the laboratory reports not provided.Device recall and injury cannot be confirmed.Root cause: the root cause of the revision surgery was pain after a fall, fracture of the trochanter, and what appeared to be trunnionosis with soft tissue metal response.The root cause of the trunnionois cannot be determined.The root cause of the other allegations cannot be determined as they cannot be confirmed.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 40MM/+4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
marisol santiago
325 corporate drive
centennial park, elstree
mahwah, NJ 07430
2018315000
MDR Report Key15534773
Report Number0002249697-2022-01429
Device Sequence Number7848310
Product Code JDI
UDI-Device Identifier07613327032741
UDI-Public07613327032741
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K061434
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2018
Device Catalogue Number6260-9-240
Device Lot NumberMME39E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/12/2022
Supplement Date Received by Manufacturer12/18/2023
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date01/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/12/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age89 YR
Patient SexMale
-
-