SMITH & NEPHEW, INC. JOURNEY FEM IMPL IMPACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 74012831
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the journey fem impl impactor is broken.It is unknown if this was noticed during surgery or not; therefore, patient involvement has not been confirmed.Upon investigation, it was found that the head of the impactor has broken from the rest of the device.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H10 h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The head of the impactor has broken from the rest of it, rendering the device inoperative.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.As the device broke and it can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY FEM IMPL IMPACTOR
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15535444
Report Number1020279-2022-04245
Device Sequence Number20772187
Product Code HRY
UDI-Device Identifier03596010556905
UDI-Public03596010556905
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74012831
Device Catalogue Number74012831
Device Lot Number14EM05960
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2022
Supplement Date Received by Manufacturer10/25/2022
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date10/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/08/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-