• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. HEALICOIL RG SA 4.75MM W/2 UB-BL CBRD BL; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. HEALICOIL RG SA 4.75MM W/2 UB-BL CBRD BL; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72203704
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a hamstring rupture procedure, two healicoil anchors were fractured.The bone holes were made using a straight awl (a dilator was not used).It is unknown how the procedure was completed or if there was a surgical delay.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H10: internal complaint reference (b)(4).H3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A visual evaluation of the device showed that it was not returned in the original packaging.Two of the sutures are still wrapped over the fork/inserter of the device and threaded through the cannula/handle.The anchor is on the inserter but missing some ribs.There is bio debris between ribs, on the sutures, and on the inserter.A functional evaluation of the device showed that the sliding ring still functioned as intended.Based on the condition of the product material found during visual inspection, additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A review of the process summary, each component should be visually inspected for all non-conforming attributes.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.This case reports, the fracture of two healicoil anchors in a hamstring rupture procedure where a straight awl instrument was reportedly used create a ¿bone holes¿.As of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.Information regarding how the procedure was completed, if back-up devices was used, or whether a procedural delay occurred are also unknown.The patient impact: per case communication, ¿no damage was caused.¿ additionally, no patient harm or injury has been alleged.Therefore, no further clinical/medical assessment is warranted.The complaint was confirmed and the root cause was associated with unintended use of the device.Factor that may have contributed to the reported event include use of excessive force during insertion can cause failure of the anchor or insertion device.No containment or corrective actions are recommended at this time.Correction in a1 (should be read in blank).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALICOIL RG SA 4.75MM W/2 UB-BL CBRD BL
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15535506
Report Number1219602-2022-01466
Device Sequence Number15411836
Product Code MAI
UDI-Device Identifier00885554030310
UDI-Public00885554030310
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151105
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/14/2025
Device Model Number72203704
Device Catalogue Number72203704
Device Lot Number2085262
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;