Model Number Z493G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown surgery on an unknown date and suture was used.During the procedure, the suture broke during suturing.It was pointed out that there may be no normal strength in the suture.The part where the suture broke was unknown.There were no adverse consequences to the patient.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent to the fda: (b)(6) 2022 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.(b)(4)device not returned.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 11/15/2022.Additional information: h6.Component code.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 11/14/2022.H3 evaluation: actual sample: the product was returned for evaluation.Visual inspection of the actual sample was conducted on the returned device.Visual analysis of the returned sample determined that it was received a winding former and a needle-suture in two sections that pertain to product code.Upon visual assessment of the needle-suture piece, it was observed that the strand had begun with a degradation process caused by exposure to the environment.In addition, the foil was not returned for evaluation to determined the assignable cause.The product code contains an absorbable suture.As the sample was received open, the time of exposure to the environment could not be determined and the functional test cannot be performed.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the reported incident.Representative sample: the product was returned for evaluation.Visual inspection of representative sample and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that it was received one unopened sample that pertain to the product code.In order to evaluate the condition of the returned samples the packet was opened, and the swage and attachment area was noted to be as expected.The suture was dispensed without problems and examined along the strand and no issues related to breakage sutures or anomalies were observed during the evaluation.A functional test was performed using equipment and the tensile force was above the minimum requirements.As part of quality process, all devices are manufactured, inspected, and released to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDSII CLR 18IN 5-0 S/A P-3 PRM MP
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez 32604
MX   32604
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15535565
Report Number2210968-2022-08166
Device Sequence Number18114721
Product Code NEW
UDI-Device Identifier10705031061156
UDI-Public10705031061156
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
N18331
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Model NumberZ493G
Device Catalogue NumberZ493G
Device Lot NumberRKMDQU
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/16/2022
Supplement Date Received by Manufacturer11/15/2022
11/15/2022
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date11/14/2022
11/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-