Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: additional product codes: hwc complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that on (b)(6) 2022, while inserting a 2.7 va locking screw, the screw did not properly engage into the locking holes of a posterior medial distal humerus plate.Additionally, when inserting a locking screw, a shaving of metal came off of the screw.It was removed and the surgery was successfully completed.No surgical delay was reported.This report involves one 2.7mm/3.5mm va-lcp postlat dhp-lat supt 4h/rt/88mm-med.This is report 1 of 3 for (b)(4).
|