• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. AIMER ENDOFEMORAL 6MM OFFSET (GREEN); ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. AIMER ENDOFEMORAL 6MM OFFSET (GREEN); ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 72201716
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during an acl procedure, when positioning the aimer endofemoral, the surgeon made a movement that broke the tip of the guide.It was completely removed from the patient with grasper forceps.It is unknown how the procedure was completed, and whether it caused or not a surgical delay.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H10: internal complaint reference:(b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that the patient impact beyond the reported events could not be determined based on the limited information provided.An analysis of the customer provided image found the broken tip detached from the guide.The root cause of the reported event could not be determined.Factors that could have contributed to the failure include an application of unintended inappropriate or excessive force to the device.No containment or corrective actions are recommended at this time.H11: corrected information in b5, h6 (health effect - impact code).
 
Event or Problem Description
It was reported that, during an acl procedure, when positioning the aimer endofemoral, the surgeon made a movement that broke the tip of the guide.It was completely removed from the patient with grasper forceps.It is unknown how the procedure was completed.No delay was reported.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIMER ENDOFEMORAL 6MM OFFSET (GREEN)
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15536401
Report Number1219602-2022-01473
Device Sequence Number14806380
Product Code NBH
UDI-Device Identifier03596010608215
UDI-Public03596010608215
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72201716
Device Catalogue Number72201716
Device Lot Number50541747
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer10/18/2023
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date10/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/25/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age26 YR
Patient SexFemale
-
-