SMITH & NEPHEW, INC. BURR 5.5MM ABRADER 180MML DISPOSABLE (3; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
| Model Number |
72200082 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on clinical study "hip arthroscopy safety and effectiveness report: accu-pass direct crescent xl, cap-fix blade, hip pack, clear trac cannula, 5.5mm abrader burr, and 4.5mm curved full radius shaver", was found that during hip capsule closure and hip labral repair procedures, a 5.5mm abrader burr was noted to be spitting metal after significant acetabuloplasty.Burr was removed and fragments were evacuated from the joint.The procedure was completed without significant surgical delay using a back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).H10 h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that was communicated via e-mail that the fragments were evacuated.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Search Alerts/Recalls
|
|
|