| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported two continuum drill bits broke.Attempts have been made and no additional information on the reported event is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: 00879000703- modular flex drill bit 3.2 mm diameter 30 mm length- 63837149.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2022 - 02864.Customer has indicated that the product will not be returned to zimmer biomet for investigation, product discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.H6 component code: mechanical (g04)- drill.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional information on the reported event.
|
| |
|