• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL BCI WW1030 PULSE OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL BCI WW1030 PULSE OXIMETER Back to Search Results
Model Number WW1030EN
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device is alarming and reading low o2 when another unit was reading normally with the same patient.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Additional Manufacturer Narrative
A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
Other, other text: h6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device in good condition.Error was found in the device ehl (event history log) multiple times.The customer stated problem was not duplicated.The monitor was powered up properly without alarm; however, when the on/off button was pressed it would not turn off.The only way to turn it off is to remover the main battery.Replaced the main board and the device powered up and turned off correctly.The main board power circuitry was shorting and it was replaced.The cause of the reported problem could not be determined.A dhr (device history review) was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during investigation.Service history review identified this device has not been in for service previously., corrected data: d5: correction: operator of the device: unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI WW1030 PULSE OXIMETER
Common Device Name
OXIMETER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15536716
Report Number3012307300-2022-20720
Device Sequence Number6972169
Product Code DQA
UDI-Device Identifier30843418000191
UDI-Public30843418000191
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K083557
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberWW1030EN
Device Catalogue NumberWW1030EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2022
Supplement Date Received by Manufacturer10/21/2022
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/04/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-