|
|
| Model Number |
X4843G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.The rope breaks during surgery.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: this hospital has received 2 boxes (24 pieces) of suture with related code and lot (code: x4843g and lot: rhmehj).During the surgery, 8 sutures have broken while knotting.Therefore, the doctor didn't want to use the rest of the sutures (16 pieces).They have discarded the 8 broken sutures (therefore they will not be returned for analysis) and have sent the 16 pieces for analysis although these 16 pieces of sutures are not used on the patient/surgery.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Did the event occur during one or multiple patient procedures? what is the total number of procedures? have any of these events been previously reported to ethicon? event related to mw # 2210968-2022-08180, mw # 2210968-2022-08181, mw # 2210968-2022-08182, mw # 2210968-2022-08183, mw # 2210968-2022-08184, mw # 2210968-2022-08186, mw # 2210968-2022-08187.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 11/2/2022.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: the incident took place during one operation to one patient.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 2/22/2023 h6 component code: g07002 no device problem found h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned devices.Visual analysis of the returned sample determined that it was received two open boxes with as total of sixteen unopened samples that pertain to the product code x4843g.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand and no issues related to breakage sutures or anomalies were observed during the evaluation.A functional test was performed using instron equipment and the tensile force was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|