| Model Number |
68LP-440 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/04/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported left side device deflation.Healthcare professional additionally reported left side" hole on patch side".Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: based on the device analysis grid, the assessments of the complaint are: deflation: observed crease smooth opening on anterior side assessed as fold flaw opening.Additional observations: crease fold observed on the device.Wear abrasion observed on the device.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side device deflation.Healthcare professional additionally reported left side" hole on patch side".Device has been explanted.
|
| |
|
Search Alerts/Recalls
|