| Model Number |
TRM-310 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
05/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported left side "hardening and exchange from textured to smooth." additionally healthcare professional reported "a left side capsular contracture baker grade iii" as well as a bilateral exchange from textured to smooth due to concern with the product".Patient further reported rupture.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation is: "hardening as well as a bilateral exchange of textured devices due to patient's concern with product.".
|
| |
|
Search Alerts/Recalls
|