• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 34CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 34CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0479-02
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event or Problem Description
It was reported that during insertion of the intra-aortic balloon (iab) the sheath could not be inserted through the guidewire because the dilator in the sheath was blocked.A new iab was inserted to continue therapy.There was no patient harm or adverse event reported.
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint (b)(4).
 
Event or Problem Description
N/a.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The product was returned with the membrane folded inside the t-handle and no blood on the exterior of the catheter.The sheath, dilator, and step dilator were returned separate from the iab.Blood was observed within the sheath.The tip of the sheath was observed to be crimped.A kink was observed on the inner lumen approximately 29.2cm from the iab tip.The guidewire, pressure tubing, and one-way valve were also returned.The one-way valve was vacuum tested and it held vacuum.The returned 0.025¿ guidewire was able to be inserted through the dilator and step dilator without difficulty.No blockage was observed.The dilator was able to be inserted into the sheath without difficulty.No blockage was observed.The reported event cannot be confirmed by the evaluation.A lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5 FR. 34CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key15538089
Report Number2248146-2022-00770
Device Sequence Number7083732
Product Code DSP
UDI-Device Identifier10607567106540
UDI-Public10607567106540
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K041281
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/09/2023
Device Model Number0684-00-0479-02
Device Catalogue Number0684-00-0474
Device Lot Number3000133536
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2022
Supplement Date Received by Manufacturer01/31/2023
03/16/2023
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date01/31/2023
03/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-