BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The patient presented with peripheral artery disease (pad).The 90% stenosed target lesion was located in the moderately tortuous and moderately calcified iliac vessel.A 8.0 x 20, 75cm mustang balloon catheter was advanced for dilation.However, during initial inflation at 10 atmospheres for 10 seconds, the balloon ruptured at the middle part of the balloon.The device was removed and the procedure was completed with a different device.There were no patient complications nor injuries reported, and the patient was in good condition after the procedure.
 
Additional Manufacturer Narrative
(b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15538338
Report Number2124215-2022-39251
Device Sequence Number15215412
Product Code FGE
UDI-Device Identifier08714729793519
UDI-Public08714729793519
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/24/2023
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0025638569
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/29/2022
Initial Report FDA Received Date10/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1