• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG EASYPUMP; PUMP, INFUSION, ELASTOMERI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG EASYPUMP; PUMP, INFUSION, ELASTOMERI Back to Search Results
Model Number 4540040-02
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: it was reported that foreign matter was observed in the fluid path past the filter of the device.No injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Several samples were provided for evaluation.The samples were visually evaluated, and foreign matter was confirmed.Although the reported defect was confirmed, the exact cause was unable to be determined at this time.A review of the batch history records was performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYPUMP
Common Device Name
PUMP, INFUSION, ELASTOMERI
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen, 34212
GM  34212
Manufacturer (Section G)
B. BRAUN MELSUNGEN AG
carl-braun strabe 1
melsungen, 34212
GM   34212
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key15538480
Report Number9610825-2022-00394
Device Sequence Number5996956
Product Code MEB
UDI-Device Identifier04046964448942
UDI-Public(01)04046964448942
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K081905
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2026
Device Model Number4540040-02
Device Catalogue Number4540040-02
Device Lot Number21H10GE672
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2022
Supplement Date Received by Manufacturer09/09/2022
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date02/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-