SMITH & NEPHEW, INC. DIAMOND RASP 90 DEG; ARTHROSCOPE Back to Search Results
Model Number 014550
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Laceration(s) (1946)
Date of Event 05/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after an arthroscopy using a diamond rasp, the patient experienced a twisting injury to the right knee.A large displaced fragment of lateral meniscus tear was noted on mri.The patient was treated by conservative management.The issue is resolving.
 
Additional Manufacturer Narrative
H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined due to the limited clinical information provided.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that the documents provided did not give any insight into the root cause of the reported injury.Although, additional details of the conservatory management that was used was requested along with images no information was provided for review.It was noted that the issue has been resolved.Based on the limited information provided a thorough medical assessment could not be performed.Based on the limited information provided we are unable to conclude on factors known to contribute to the alleged report.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMOND RASP 90 DEG
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15538601
Report Number1219602-2022-01480
Device Sequence Number20594606
Product Code HRX
UDI-Device Identifier03596010096111
UDI-Public03596010096111
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K932460
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number014550
Device Catalogue Number014550
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2022
Supplement Date Received by Manufacturer02/15/2024
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date02/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-