| Model Number |
32375 |
| Medical Device Problem Codes |
Break (1069); Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter city - (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter broke.A flexima nephrostomy drainage catheter had been placed for drainage of the kidney.It was noted there was difficulty pulling the suture and the catheter broke.A new flexima catheter was used.There were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter broke.A flexima nephrostomy drainage catheter had been placed for drainage of the kidney.It was noted there was difficulty pulling the suture and the catheter broke.A new flexima catheter was used.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter city - (b)(6).Device analysis by manufacturer: unit returned with its original pouch.The catheter was returned for analysis; moreover, the flexible cannula was loaded in the catheter, and no damages were observed.The flexible cannula was inserted and retracted from the catheter without issues.
|
| |
|