• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIPRO (THAILAND) CORP. LTD. NIPRO BLOOD TUBING SET W/TP; BLOODLINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIPRO (THAILAND) CORP. LTD. NIPRO BLOOD TUBING SET W/TP; BLOODLINE Back to Search Results
Model Number BL+A209Y/V803
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
Rn checked machine due to alarm, noted air in the lines, checked to see if chambers were low and they were not, moved over lines to check for faults and blood was noted coming out of the pump.Lines were clamped and no blood was returned due to a tear in the lines.Patient was stable bp 136/58, pulse 54.Patient was told that there was a problem with the lines and they needed to be changed, patient was clam.Lines were taken off the machine, double bagged in biohazard bags with original packaging.Machine was cleaned and newlines were primed, treatment restarted w/o further incident.Estimated 400 cc blood loss, md aware, v/s 155/64 pulse 54, patient alert and oriented w/o complaints closely monitor for any blood infection draw cbc on sat.
 
Event or Problem Description
Rn checked machine due to alarm, noted air in the lines, checked to see if chambers were low and they were not, moved over lines to check for faults and blood was noted coming out of the pump.Lines were clamped and no blood was returned due to a tear in the lines.Patient was stable bp 136/58, pulse 54.Patient was told that there was a problem with the lines and they needed to be changed, patient was clam.Lines were taken off the machine, double bagged in biohazard bags with original packaging.Machine was cleaned and newlines were primed, treatment restarted w/o further incident.Estimated 400 cc blood loss, md aware, v/s 155/64 pulse 54, patient alert and oriented w/o complaints closely monitor for any blood infection draw cbc on sat.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIPRO BLOOD TUBING SET W/TP
Common Device Name
BLOODLINE
Manufacturer (Section D)
NIPRO (THAILAND) CORP. LTD.
10/2 moo 8
bangnomko,
sena, ayuthaya 13110 ,
TH  13110,
Manufacturer (Section G)
NIPRO (THAILAND) CORP. LTD.
10/2 moo 8
bangnomko,
sena, ayuthaya 13110 ,
TH   13110,
Manufacturer Contact
michelle tejada
3150 nw 107th avenue
miami, FL 33172
3055997174
MDR Report Key15538875
Report Number8041145-2022-00010
Device Sequence Number7046498
Product Code FJK
UDI-Device Identifier00383790004053
UDI-Public00383790004053
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K072024
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBL+A209Y/V803
Device Catalogue NumberBL+A209Y/V803
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/14/2022
Supplement Date Received by Manufacturer09/14/2022
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date01/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-