Blank fields on this form indicate the information is unchanged, unknown, or unavailable.Investigation ¿ evaluation a representative of (b)(6) hospital (taiwan) informed cook that on (b)(6) 2022 the blue port in a cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter set (rpn: c-utlm-701j-rsc-abrm-hc-rd; lot: 14322788) was occluded.The occlusion was only noted after the procedure and was replaced with another catheter right away.No medications had been infused through the device yet.It was reported that there were no adverse effects to the patient due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, and quality control procedures, as well as a visual inspection and functional test of the returned device, were conducted during the investigation.Cook received one used c-utlm-701j-rsc-abrm-hc-rd catheter.After tabletop testing, the obstruction was removed and tested.The test found it was blood/biomatter that was obstructing the lumen.Additionally, a document-based investigation evaluation was performed.In response to this incident, cook completed a review of the product device master record (dmr) and concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Cook also reviewed the device history record (dhr).The dhr for lot 14322788 records no non-conformances.A search on the sub assembly lots and raw material lots found one relevant non-conformance.There are inspections in place to identify this failure, and the non-conforming product was scrapped.Since there are no related non-conformances or other complaints on this lot, cook concluded that no non-conforming product from this lot exists in house or in the field.Cook also reviewed product labeling.The product ifu, [c_t_ctulmabrm_rev7] ¿cook spectrum central venous catheter minocycline/rifampin antibiotic impregnated power injectable,¿ provides the following information to the user related to the reported failure mode: warnings ¿failure to ensure patency of the catheter prior to power injection studies may result in catheter failure.¿ clincal studies product recommendations suggested catheter maintenance ¿to prevent clotting or possibility of air embolus, the double-lumen¿s #2 lumen, the triple-lumen¿s #2 and #3 lumens and the five-lumen¿s #2, #3, #4, and #5 lumens should be filled with saline solution or heparinized saline solution (100 units of heparin per ml of saline is usually adequate), depending on institutional protocol, prior to catheter introduction.¿ instructions for use ¿2.Prepare the catheter for insertion by flushing each of the lumens and clamping or attaching the injection caps to the appropriate extensions.¿9.-lumens should now be flushed with 5-10ml normal saline prior to use or establishment of heparin lock.¿ the information provided upon review of device master record, product labeling, device history record, and device failure analysis, does not indicate the device was manufactured out of specification.There is no evidence of nonconforming material in house or in the field.Based on the information provided, inspection of the returned device, and the results of the investigation, cook concludes the main cause of this event is related to the procedure.During the device failure analysis, it was determined that the device had been occluded with biomatter/blood.It is possible that the device wasn¿t flushed before use, which could have contributed to the occlusion.Cook is unable to confirm this without more information.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|