• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA POSTERIOR STABILIZED ARTICULAR SURFACE RIGHT 10MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA POSTERIOR STABILIZED ARTICULAR SURFACE RIGHT 10MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during knee arthroplasty when the surgeon was inserting the articular surface, the implant could not be seated on the tibial tray after several attempts.The procedure was completed with another articular surface of the same size.No adverse events were reported as a result of this malfunction.
 
Additional Manufacturer Narrative
(b)(4).Report source: foreign: italy.The complainant has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual evaluation of the returned device identified that the dovetail feature was flared out and compressed.The device also exhibited damage on other areas.The device history records were reviewed and no discrepancies were identified.Damage to the articular surface can occur if the articular surface is not correctly placed and oriented prior to pushing it in with the inserter.Detailed instructions for insertion of the articular surface are provided within the surgical technique.Based on the evaluation performed and information received, the root cause is attributed to use error.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA POSTERIOR STABILIZED ARTICULAR SURFACE RIGHT 10MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15539448
Report Number3007963827-2022-00250
Device Sequence Number7083781
Product Code JWH
UDI-Device Identifier00889024643512
UDI-Public(01)00889024643512(17)270531(10)65376585
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42521401010
Device Lot Number65376585
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2022
Supplement Date Received by Manufacturer02/07/2023
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date02/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-