CETERIX ORTHOPAEDICS, INC NOVOCUT SUTURE MANAGER; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number CTX-C001
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Laceration(s) (1946)
Date of Event 05/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after an arthroscopy using a novocut suture, the patient experienced a twisting injury to the right knee.A large displaced fragment of lateral meniscus tear was noted on mri.The patient was treated by conservative management.The issue is resolving.
 
Additional Manufacturer Narrative
H6, health effect - impact code was added.H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Based on the limited information provided a thorough medical assessment could not be performed.There was no relationship found between the device and the reported event.The complaint was not confirmed.Factors that could have contributed to the reported event include a failure of a concomitant device.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVOCUT SUTURE MANAGER
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
CETERIX ORTHOPAEDICS, INC
6500 kaiser dr ste 120
fremont CA 94555
Manufacturer (Section G)
CETERIX ORTHOPAEDICS, INC
6500 kaiser dr ste 120
fremont CA 94555
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15540073
Report Number3009131204-2022-00021
Device Sequence Number1855866
Product Code NBH
UDI-Device Identifier00853541006129
UDI-Public00853541006129
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberCTX-C001
Device Catalogue NumberCTX-C001
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2022
Supplement Date Received by Manufacturer01/11/2023
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-