It was reported that, after an arthroscopy using a novocut suture, the patient experienced a twisting injury to the right knee.A large displaced fragment of lateral meniscus tear was noted on mri.The patient was treated by conservative management.The issue is resolving.
|
|
H6, health effect - impact code was added.H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Based on the limited information provided a thorough medical assessment could not be performed.There was no relationship found between the device and the reported event.The complaint was not confirmed.Factors that could have contributed to the reported event include a failure of a concomitant device.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|