EXACTECH, INC. ALT HA S CLR STD SZ 5; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU
|
Back to Search Results |
|
| Model Number |
ALT HA S CLR STD SZ 5 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Implant Pain (4561)
|
| Date of Event |
09/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported a female patient, initial right hip implanted on (b)(6) 2017, was brought back for right hip pain and was revised on (b)(6) 2022, approximately 4 years 10 months post the initial implant procedure.The femur was dislocated, the femoral head removed.The stem was checked and found to be loose and removed.Attention turned to the cup, the poly was removed with the help of a chisel.The cup was checked and found to be well fixed.A new liner was implanted.Attention turned back to femur where a size 8 alteon hps stem was implanted along with a new biolox head.Patient was last known to be in stable condition following the event.No other patient information/medical history provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitants: 130-32-51 - nv gxl liner neutral, 32mm id group 1 cups serial : (b)(4).142-32-00 - cocr fem head 32mm +0 offset 12/14 serial: (b)(4).186-01-48 - integrip cc, cluster 48mm, g1 serial: (b)(4).Z-2117-2021.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h4, h6.Mdr section codes updated/corrected: b, c, d, g.The revision reported may have been the result of femoral stem loosening as reported, however this cannot be confirmed as pre-revision radiographs and relevant clinical information were not provided.User and patient-related considerations could not be evaluated from the information provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|