| Catalog Number |
30873 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd alaris¿ pca asv microbore cv the product was damaged.This occurred twice.There was no report of patient impact.The following information was provided by the initial reporter: we recently had 2 sets of bd pca administration sets break.Both breaks happened to the same patient at the same place in the same 24 hours.
|
| |
|
Additional Manufacturer Narrative
|
|
Medical device expiration date: unknown.Device manufacture date: unknown.A complaint of two sets breaking was received from the customer.No product or photo was returned by the customer.The customer complaint of component damage could not be verified due to the product not being returned for failure investigation.A device history record review could not be performed on model 30873 because the lot number is unknown.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: a complaint of two sets breaking was received from the customer.No product or photo was returned by the customer.The customer complaint of component damage could not be verified due to the product not being returned for failure investigation.A device history record review could not be performed on model 30873 because the lot number is unknown.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd alaris¿ pca asv microbore cv the product was damaged.This occurred twice.There was no report of patient impact.The following information was provided by the initial reporter: we recently had 2 sets of bd pca administration sets break.Both breaks happened to the same patient at the same place on the same 24 hours.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 19-jan-2023.H6: investigation summary a complaint of two administration sets breaking was received from the customer.Two samples of material #30873 were returned for investigation.Through visual inspection, the customer complaint was confirmed.The female luers were cracked.When a syringe was connected, the sets would leak from the connection.A device history record review could not be performed on model 30873 because the lot number is unknown.The manufacturing plant was notified of this defect.An investigation was performed, and it was determined that there were two possible root causes for this issue, use of alcohol on the female luer or a supplier issue.The complaint was escalated to the supplier and samples were sent for investigation.The samples were inspected by the supplier and no evident flow lines or signs on the parts that justify weakness due to the molding process.The samples were also measured with a calibrated gauge conforming to iso standard and the female luers are within specification.Due to the lot being unknown, a device history review was unable to be performed by the supplier to determine if there were any issues during the molding of this component, or which cavity was used for this lot.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd alaris¿ pca asv microbore cv the product was damaged.This occurred twice.There was no report of patient impact.The following information was provided by the initial reporter: we recently had 2 sets of bd pca administration sets break.Both breaks happened to the same patient at the same place on the same 24 hours.
|
| |
|
Search Alerts/Recalls
|