| Catalog Number |
138-32-51 |
| Medical Device Problem Codes |
Fracture (1260); Device Dislodged or Dislocated (2923); Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Fall (1848); Failure of Implant (1924); Muscle Weakness/Atrophy (1967); Pain (1994); Discomfort (2330); Joint Dislocation (2374); Ambulation or Postural Difficulties (2544); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported via a legal notification that a female patient, initial right hip connexion gxl acetabular liner implanted on (b)(6) 2016, underwent a revision procedure on (b)(6) 2019, approximately 3 years 1 month post the initial implant procedure for reasons unknown.No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
No device return anticipated due to this being a legal case.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The cause of the subsequent revision cannot be conclusively determined, insufficient information.
|
| |
|
Additional Manufacturer Narrative
|
|
D10.Concomitants - product information: 4272018 170-32-50 - biolox delta option femoral head 32mm od; 4374069 170-50-00 - biolox delta adapter 16/18-12/14; +0mm.There is no other information available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10.Investigation result: the most likely cause for the revision reported due to prosthesis wear is a combination of risk factors including, use error, implant positioning, implant size selection, and patient factors may have also been a contributing factor to the early prosthesis wear.However, this is not confirmed, the devices were not returned.These devices used in treatment and not for diagnosis.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b5, e1, e2, e3, g1, h6 - health effect - clinical code, health effect - impact code, medical device problem code, component code, type of investigation, investigation findings and h11.The gxl acetabular liner involved was reportedly revised due to prosthesis wear, approximately 3 years and 3 months after the index surgery.However, the revised components were not returned to exactech for evaluation and no images or radiographs were provided.Based on the available information, the patient involved meets the following risk criteria for early prosthesis wear as specified in the hhe: combination of largest available femoral head and/or thinnest available acetabular liner was used, patients implanted with a lateralized liner.The most likely underlying cause for the revision due to early prosthesis wear.However, this cannot be confirmed from the reported information and the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial left total hip arthroplasty.Subsequently, the patient experienced pain, ambulation difficulties and a fall.As a result, the patient underwent a left hip revision approximately 3 years and 2 months after initial implantation.During the revision it was noted that the hip had dislocated and the acetabular liner was worn and fractured.Patient was last known to be in stable condition.No further information available.
|
| |
|