• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL SAFIRE¿ BI-DIRECTIONAL ABLATION CATHETER 4 MM TIP UNIVERSAL TEMPERATURE MONITORI; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL SAFIRE¿ BI-DIRECTIONAL ABLATION CATHETER 4 MM TIP UNIVERSAL TEMPERATURE MONITORI; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number 402810
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Heart Block (4444)
Date of Event 09/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
During an atrioventricular nodal re-entry procedure a heart block occurred.During one of the rf applications, the ablation catheter was noted to have risen into the fast pathway area and heart block was observed.Conduction quickly recovered and another rf application was applied to the slow pathway location.After the application, the catheter swung through the his zone and heart block was once again observed.This time 2:1 heart block that recovered to mobitz type 1 and then it recovered to 1:1.The patient was given a temporary pacing wire and sent to the floor for observation and then was given a permanent pacemaker implant a few days later.
 
Additional Manufacturer Narrative
The reported event of "heart block" could not be confirmed.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFIRE¿ BI-DIRECTIONAL ABLATION CATHETER 4 MM TIP UNIVERSAL TEMPERATURE MONITORI
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345
Manufacturer (Section G)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15541387
Report Number2182269-2022-00042
Device Sequence Number12171278
Product Code LPB
UDI-Device Identifier05414734206013
UDI-Public05414734206013
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P960016-S014
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model Number402810
Device Catalogue Number402810
Device Lot Number8450780
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date10/04/2022
Supplement Report FDA Received Date11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
Patient SexMale
-
-