ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Medical Device Problem Code Visual Prompts will not Clear (2281)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported, the autopulse platform (sn (b)(4)) displayed "system error, out of service, revert to manual cpr" upon powering up.The customer rebooted the platform and replaced the battery but the issue was not resolved.It is unknown when the problem occurred.However, patient use information was requested but no additional information was provided.
 
Additional Manufacturer Narrative
The reported complaint of the autopulse platform (sn (b)(4)) displayed the "system error, out of service, revert to manual cpr" error message was confirmed during initial functional testing and archive data review.The root cause for the system error was due to the defective processor board, as a result of the device's aging.The autopulse platform is a reusable device that was manufactured in october 2006 and is 16 years old, well beyond the expected service life of 5 years.During visual inspection, there was no physical damage observed on the autopulse platform.The archive data was corrupted due to the defective processor board, however, the system error 136 (internal parameter corrupted) error message was noted in the archive, thus confirming the reported complaint.During the initial functional test, the platform displayed a "system error, out of service, revert to manual cpr" error message during power on, thus confirming the reported complaint.To remedy the error message, the defective processor board needs to be replaced.The last autopulse 1.1 was manufactured in 2009.As of july 2016, zoll announced the end of life for autopulse 1.1.Many important components used in autopulse 1.1 are no longer available, further restricting our ability to service.The platform will be returned to the customer without the repair and it will have a label stating "not for clinical use" or will be scrapped.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaints reported for autopulse platform with serial number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Common Device Name
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave,
san jose, CA 95131
MDR Report Key15541555
Report Number3010617000-2022-01721
Device Sequence Number6501800
Product Code DRM
UDI-Device Identifier00849111000277
UDI-Public00849111000277
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K112998
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-01-66
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2022
Initial Report FDA Received Date10/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2006
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1