• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH 32MM STD LFIT V40 HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH 32MM STD LFIT V40 HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 6260-9-132
Medical Device Problem Codes Corroded (1131); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Metal Related Pathology (4530)
Date of Event 06/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on her left hip on or about (b)(6) 2011 and was revised on (b)(6) 2022.It is further alleged that she suffered injuries as a result of implantation and explantation of the device at issue, device recall and excessive levels of chromium and cobalt in her blood.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on her left hip on or about (b)(6) 2011 and was revised on (b)(6) 2022.It is further alleged that she suffered injuries as a result of implantation and explantation of the device at issue, device recall and excessive levels of chromium and cobalt in her blood.
 
Additional Manufacturer Narrative
Corrected data: manufacturing site for devices.An event regarding corrosion and abnormal ion level involving a metal head was reported.The event of corrosion was confirmed.Abnormal ion level was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical information by a clinical consultant indicated: conclusion/assessment: the patient underwent a routine total hip arthroplasty in 2011.No medical records were provided for the preoperative or postoperative course.She apparently underwent revision hip arthroplasty in 2022.Again, no medical records, preoperative evaluation, radiographs or postoperative care records were provided for review.At the time of revision surgery.There were a small amount of corrosive products noted on the trunnion that were wiped free and the head and liner were exchanged.Increased metal levels were noted in the operative report but not provided for review.No laboratory studies were provided for review.This case is apparently evolved into a legal filing.Event confirmation: a primary total left hip arthroplasty in 2011 can be confirmed.A revision left hip arthroplasty in 2022 can be confirmed.Increased metal levels cannot be confirmed.Injury cannot be confirmed.Device recall cannot be confirmed.Root cause: the root cause for the total hip arthroplasty in 2011 was hip arthritis.The root cause for the revision hip arthroplasty in 2022 was stated as failed left hip arthroplasty secondary to trunnion corrosion, however corrosion products, increased metal levels, or the reasons for failure cannot be confirmed as a root cause cannot be ascertained.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of the provided medical information by a clinical consultant indicated: conclusion/assessment: the patient underwent a routine total hip arthroplasty in 2011.No medical records were provided for the preoperative or postoperative course.She apparently underwent revision hip arthroplasty in 2022.Again, no medical records, preoperative evaluation, radiographs or postoperative care records were provided for review.At the time of revision surgery.There were a small amount of corrosive products noted on the trunnion that were wiped free and the head and liner were exchanged.Increased metal levels were noted in the operative report but not provided for review.No laboratory studies were provided for review.This case is apparently evolved into a legal filing.Event confirmation: a primary total left hip arthroplasty in 2011 can be confirmed.A revision left hip arthroplasty in 2022 can be confirmed.Increased metal levels cannot be confirmed.Injury cannot be confirmed.Device recall cannot be confirmed.Root cause: the root cause for the total hip arthroplasty in 2011 was hip arthritis.The root cause for the revision hip arthroplasty in 2022 was stated as failed left hip arthroplasty secondary to trunnion corrosion, however corrosion products, increased metal levels, or the reasons for failure cannot be confirmed as a root cause cannot be ascertained.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32MM STD LFIT V40 HEAD
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NJ NA
EI   NA
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15542057
Report Number0002249697-2022-01432
Device Sequence Number20479327
Product Code JDI
UDI-Device Identifier07613327013047
UDI-Public07613327013047
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/29/2016
Device Catalogue Number6260-9-132
Device Lot Number36033108
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/12/2022
Supplement Date Received by Manufacturer01/02/2024
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date01/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/24/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age75 YR
Patient SexFemale
-
-