• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH ENTFLX 12FR 43IN W STYL YPORT; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH ENTFLX 12FR 43IN W STYL YPORT; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884721252
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907); Material Twisted/Bent (2981)
Health Effect - Clinical Code Laceration(s) (1946)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported they inserted the tube and an x-ray confirmed that it was fit for use.When infusing medication, there was resistance.They verified bending of the material during its passage and removal of the mandrel (guide wire).The mandrel had extreme resistance when being removed and the green connector disconnected from the wire, hurting hands and cutting gloves.
 
Additional Manufacturer Narrative
Additional information: h4 device manufacturer date was added.The device history record (dhr) review showed no abnormal process conditions were present during the manufacturing of the product that could have led to the reported condition.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.No sample evaluation could be performed because no sample nor photo was received for evaluation.The reported condition could not be confirmed.The investigation was carried out with the multifunctional team.All processes and controls were found to be followed correctly, including packaging and all inspections performed on the product.No abnormal conditions were found that could trigger the reported condition.Current controls include a 100% visual inspection by production personnel according to standard work instructions.In addition, a quality sampling inspection per procedure, is performed to ensure that product is free from missing components, damaged components, inappropriate assembly, and contamination.Current process and controls were found to be followed correctly, including subassemblies, finished product assembly, packaging, and product inspections.No abnormal conditions were found that could trigger the reported condition.Based on the analysis, no action plan is required at this point.The manufacturing site will continue to monitor customer complaints and feedback notifications for adverse trends that require immediate attention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTFLX 12FR 43IN W STYL YPORT
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15542528
Report Number9612030-2022-03413
Device Sequence Number15080812
Product Code KNT
UDI-Device Identifier10884521518919
UDI-Public10884521518919
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8884721252
Device Catalogue Number8884721252
Device Lot Number2204101164
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2022
Supplement Date Received by Manufacturer09/23/2022
Initial Report FDA Received Date10/05/2022
Supplement Report FDA Received Date11/04/2022
Date Device Manufactured02/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-