| Model Number |
8884715122 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported a family member of a male patient under 1 year of age who is hospitalized by the infants service with a medical diagnosis of mixed chronic malnutrition, neurogenic dysphagia and dilatated heart disease who uses a 12fr gastrostomy tube called stating that the tube has a leak.The customer further stated that it was verified; breakage of the union of the key is evidenced.No patient injury was reported.
|
| |
|
Event or Problem Description
|
|
Added the actual initial reporter and moved jose salcedo to the sales rep contact.Added the no sample rationale.Added clarification to the description summaries.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on additional information received on 05oct2022, the following data was corrected: b5 describe event or problem (added additional information); e1 initial reporter name (added the correct initial reporter's first and last name); and h6 evaluation codes type of investigation (added the code for device discarded).
|
| |
|
Additional Manufacturer Narrative
|
|
Samples were not received for the investigation.The device history record was reviewed and indicated that the product was released accomplishing all quality standards.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the reported issue and determine a root cause.However, a corrective and preventive action has been initiated to address the reported issue.
|
| |
|